A certificate of analysis (CoA) is the testing record for one batch of a material. For anyone sourcing research peptides, it is the main evidence that a vial holds the right compound at the stated purity. Many CoAs look official and say little. This guide walks through a typical peptide CoA section by section, explains what each result measures, and lists the gaps worth questioning.
What a CoA covers
A CoA applies to one lot, sometimes called a batch. A report for a different lot tells you nothing about the material in your vial, even if the compound name is the same. The report states which tests were run and what they found. It is a record of results, and it says nothing about safety or fitness for any use.
CoAs come from two kinds of source. A manufacturer's in-house quality lab tests its own output. An independent third-party lab tests samples sent by a manufacturer, distributor or buyer. Both are common. A third-party report adds independence, provided you can confirm the lab issued it.
The header: identifying the material
The top of the report identifies the sample. Expect most of these fields:
- Product name
- Amino acid sequence, in one-letter or three-letter code
- Molecular formula and theoretical molecular weight
- Lot or batch number
- Manufacture date and test date
- Report number
- Name and address of the issuing lab
Two checks matter most here. The lot number on the report should match the lot number on your vial exactly. The dates should run in a sensible order: manufacture, then testing, then sale.
Identity: mass spectrometry
Mass spectrometry (MS) answers the question "is this the right molecule?" It measures the mass of the molecules in the sample. Electrospray ionization (ESI) and MALDI-TOF are the two methods you will see most often.
A useful MS result lists two numbers: the theoretical mass calculated from the sequence, and the observed mass from the instrument. They should agree within the tolerance the lab states. A spectrum image should accompany them.
ESI spectra often show the peptide carrying more than one charge. A peak labeled [M+2H]2+, for example, appears at roughly half the molecular mass. A good report labels these peaks or gives a single deconvoluted mass, so you don't have to work it out.
HPLC alone can't confirm identity. A sample can produce one clean peak and still be the wrong compound. Identity needs a mass result.
Purity: HPLC
Reverse-phase high-performance liquid chromatography (RP-HPLC) answers a different question: "how much of the sample is the main component?" The instrument separates the sample into components. A UV detector records each one as a peak, usually at 214 or 220 nm, where peptide bonds absorb light.
Purity is the area of the main peak as a percentage of the total peak area. A good report shows:
- The chromatogram image
- An integration table with each peak's retention time and area
- Method details: column type, gradient and detection wavelength
What a purity figure leaves out
HPLC purity only counts what the detector sees at that wavelength. It does not measure water, salts or counter-ions, which make up part of the weight of most lyophilized peptides. Impurities that elute under the main peak are counted as main peak. A figure of exactly 100% deserves a question, since real chromatograms almost always show small secondary peaks.
Net peptide content
A vial of lyophilized peptide holds the peptide plus counter-ions and some residual water. Net peptide content is the share of the weighed material that is peptide. Labs measure it by amino acid analysis or by nitrogen analysis.
Net peptide content is often well below 100%, and it is a separate number from HPLC purity. A sample can be highly pure and still have a modest net peptide content, because the purity figure ignores counter-ions and water. If a vial says "10 mg", ask whether that is the gross weight of material or the net amount of peptide. Our guide to salt forms and net peptide content covers this in more detail.
Other tests you may see
Many CoAs stop at identity and purity. Some add further tests, depending on the application:
- Water content, often by Karl Fischer titration
- Counter-ion content, such as TFA or acetate, by ion chromatography
- Appearance, for example "white lyophilized powder"
- Solubility in a stated solvent
- Endotoxin, by the LAL method
- Residual solvents
Which of these matter depends on the research. Cell-based work, for example, may call for counter-ion and endotoxin data that a simple chemistry application does not need.
Signs a CoA needs a closer look
- No lot number, or a lot number that doesn't match the vial.
- No issuing lab named, or no way to contact it.
- A purity figure with no chromatogram.
- Identity marked "conforms" or "passed" with no mass value.
- The same chromatogram or spectrum image appearing on reports for different compounds.
- Dates out of order, or a test date years before the sale.
- Fonts, alignment or numbers that look edited.
- A client name that differs from the seller. This can happen legitimately when material is resold, so ask.
How to verify a CoA
A report is only as good as your ability to confirm it. Three steps cover most cases.
- Contact the lab. Send the report number and ask whether the lab issued it and whether the results match its records. Some labs offer an online lookup by report number.
- Check the lab's accreditation. ISO/IEC 17025 is the international standard for the competence of testing and calibration labs. Accreditation bodies publish directories of accredited labs and the scope of methods each one is accredited for. Check that the scope covers the methods on the report. Our verification directory links to accreditation resources.
- Retest if it matters. For important work, send a sealed sample from your own lot to an accredited lab and compare the results with the supplier's report.
A quick CoA checklist
- Lot number matches the vial.
- Issuing lab is named and reachable.
- MS result shows theoretical and observed mass, with a spectrum.
- HPLC result shows a chromatogram, an integration table and the method.
- Net peptide content is stated, or the supplier can tell you.
- Dates are in order and recent.
- The lab confirms the report.
For the words that appear on CoAs, see the glossary. For warning signs beyond the paperwork, see our supplier red flags checklist.